GDSF-311WPD 3-Ni-1 SF6 Gas Oluyanju
GDSF-311WPD (GDSF-411WPD) jẹ ohun elo pipe nigbati o jẹ dandan lati ṣe idanwo akoonu omi, mimọ ati awọn ọja jijẹ ti gaasi SF6.Ẹya ipilẹ ti idanwo aaye ìri jẹ awọn sensọ jara DRYCAP® ti a ṣe nipasẹ ile-iṣẹ Finland Vaisala.Ofin infurarẹẹdi SF6 sensọ idanwo mimọ gaasi ti iṣelọpọ nipasẹ Ile-iṣẹ sensọ Yuroopu ni a lo lati ṣe idanwo mimọ ti gaasi SF6.Sensọ elekitirokemika ti Ile-iṣẹ Mombabo ṣe ni Switzerland ni a lo bi eroja oye fun idanwo awọn ọja jijẹ gaasi SF6.Pẹlu awọn eerun ohun elo amọdaju ati awọn algoridimu sọfitiwia ti o dara julọ ti STMicroelectronics, a ti ṣe agbejade iran tuntun ti olutupa gaasi 3-in-1.
SF6Ọriniinitutu
●Ọna wiwọn: Resistive ati Capacitive odiwọn opo
●Iwọn wiwọn: aaye ìri -80℃--+20℃(ppmv atilẹyin)
●Yiye: ± 1℃
(nigbati iwọn otutu aaye ìri ba wa ni isalẹ 0 ℃, abajade sensọ jẹ aaye Frost)
●Akoko Idahun: 63% [90%]
+20→-20℃ Td 5s[45s]
-20→-60℃ Td 10s[240s]
●Ipinnu: 0.01 ℃
●Atunṣe: ± 0.5 ℃
●Ẹka ifihan: ℃, ppm, ℃P20 (iye iyipada ni 20℃)
●Gaasi sisan: 400-600ml / min
●Ifihan ṣiṣan: 0-1000mL mita ṣiṣan oni-nọmba
●Ayẹwo gaasi titẹ: ≤1MPa
SF6Mimo
●Ọna wiwọn: Ilana wiwọn infurarẹẹdi (awọn sensọ jara NDIR)
●Iwọn wiwọn: 0 ~ 100% SF6
●Akoko Idahun: [90%] 60s
●Yiye: ± 0.5% FS
●Atunṣe: ± 0.5%
●Ipinnu: 0.01%
●Ẹka ifihan:%
SF6Awọn ọja ibajẹ
●Ọna wiwọn: Ilana Awọn wiwọn elekitirokemika (Awọn sensọ jara elekitiroki)
●Iwọn wiwọn: H2S: 0 ~ 200ppmv
SO2: 0 ~ 200ppmv
CO: 0 ~ 500ppmv
HF: 0 ~ 50ppmv
●Yiye: H2S: 10ppmv ± 3%
SO2: 10ppmv ± 3%
CO: 50ppmv ± 4%
HF: 10ppmv ± 3%
●Atunṣe: H2S: 10ppmv ± 2%
SO2: 10ppmv ± 2%
CO: 50ppmv ± 2%
HF: 10ppmv ± 2%
●Ipinnu: 0.01ppmv
●Ẹka ifihan: ppmv
●Ipese agbara: 220VAC ± 10%, 50Hz, AC / DC lilo, idaabobo gbigba agbara, iṣẹ-ṣiṣe ti nlọ lọwọ ko kere ju 8hours.
●Lo iwọn otutu ayika: -20--+60 ℃
●Ọriniinitutu ayika: 90% RH
●Ipa iye iwọn: Ko si ipa ti titẹ ati sisan
●Iwọn: 395 * 295 * 155mm
●Iwọn: nipa 2kg.
GDSF-411WPD SF6Awọn ọja ibajẹ
●Yiye: H2S: <10ppmv ± 0.3ppmv;> 10ppmv ± 3%
SO2: <10ppmv ± 0.3PPmv;> 10ppmv ± 3%
●Atunṣe: H2S: <10ppmv ± 0.2ppmv;> 10ppmv ± 2%
SO2: <10ppmv ± 0.2PPmv;> 10ppmv ± 2%
●Atunlo ati iṣẹ kikun: iyẹwu atunlo 2L,akoko atunlo ≤ 5min
Pẹlu gaasi gbigbe iṣẹ
●Iwọn: 395 * 295 * 315mm
●Iwọn: nipa 5kg.
Oludanwo akọkọ | 1 nkan |
Adapter | 1 ṣeto |
paipu Teflon (pẹlu àtọwọdá atunṣe sisan ati asopo iyara) | 1 ṣeto |
paipu iru | 1 ṣeto |
Awọn ohun elo | 1 ṣeto |
Ṣaja | 1 nkan |
Olumulo's Itọsọna | 1 nkan |
Iroyin igbeyewo | 1 nkan |